Gel Permeation Chromatography (GPC), otherwise known as Size Exclusion Chromatography (SEC), is a commonly used technique for determination of molecular weight. In GPC/SEC, a dilute solution of the analyte in question is prepared using a good solvent for the analyte to ensure good dissolution and expansion of the polymer chains in solution. The solution is injected into a packed column containing a distribution of pore sizes which separate the polymer chains by size. Exiting the column, the size-separated chains are effectively ‘counted’ by a refractive index detector according to their elution time from the column.
A series of calibration standards (polymers with known molecular weight) are also run through the column, which allows correlation of the elution time to a molecular weight. This analysis allows the calculation of a molecular weight distribution, along with molecular weight moments (number-, weight-, viscosity-, z-averaged, and z+1-averaged molecular weights).
In addition to the standard refractive index detector, CPG can also perform multi-angle light scattering to provide absolute molecular weight, and viscometry, to provide a universal calibration method. The coupling of light scattering with viscometry, termed triple detection GPC, also allows determination of degree of branching in branched polymers.
Inside our Agilent PL-GPC-220's oven.
Successful GPC/SEC characterization relies on understanding the proper solvent conditions, temperature, concentration, and column setup, along with knowledgeable interpretation of the results.
CPG can perform GPC on water soluble polymers natural materials, such as hyaluronic acid, as well as organic-soluble materials, even at elevated temperatures. Some examples of polymers that CPG has tested:
GPC/SEC is run to help determine physical property differences between samples, such as tensile strength or wear behavior, as well as processing differences, as melt viscosity of polymers can scale with molecular weight raised to the factor of 3 or more. GPC is particularly useful in assessing degradation behavior in materials, as well as lot-to-lot variation in raw resin.
ASTM D6474 Standard Test Method for Determining Molecular Weight Distribution and Molecular Weight Averages of Polyolefins by High Temperature Gel Permeation Chromatography
ASTM D5296 Standard Test Method for Molecular Weight Averages and Molecular Weight Distribution of Polystyrene by High Performance Size-Exclusion Chromatography
ISO 13885-1 Binders for Paints and Varnishes -- Gel Permeation Chromatography (GPC) -- Part 1: Tetrahydrofuran (THF) as Eluent
Why use Cambridge Polymer Group as your contract GPC testing lab?
Contact us for your free GPC quote.